87

鉛離子選擇電極


專門測量濃度為0.2 ppm至20,700 ppm(1x10-6 M至0.1 M)的Pb + 2。 應用包括(不限於):電鍍浴。


Model # PB21502
Slope 25 ± 3 mV between 10-100 ppm @ 25°C
Reproducibility ± 2%
Interferences Ag+1, Hg+2, Cu+2, high levels of Cd+2, Fe+2
Temperature Range 0 to 80° C (if calibrated at that temperature)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry
Short term, store in dilute lead standard
Concentration Range 0.2 ppm to 20,700 ppm  (1 x 10-6 M to 0.1 M)
pH Range 3 to 8 pH
Resistance < 1 Megaohm
Care, Maintenance & Cleaning Polish with polishing strips
Treat with a dilute ISA
Isopotential Point 50,000 ppm
On-Line Capability Yes
Applications Plating baths